<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>26007</ErrorNumber>
<ErrorFamily>Operator</ErrorFamily>
<ErrorMessage>
OPT: Timed out waiting for access to operator prompts.
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The firmware has behavioral problems if the troubled area can be worked around
great, otherwise consider another version of firmware.  A log file of the
occurance sent to the RC or firmware would be appreciated.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Resume if option given otherwise a reset will likely be necessary.
</text></DescriptionText>
</OperatorAction>
<Cause>
Somewhere the pool of prompts has been reserved and the operations being done
have taken longer than 1 second, causing another potential user to be blocked
from using the prompt pool.
</Cause>
<SeeAlso>
<UnnumberedList>
None.
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:34:12 2008 PST
</Updated>
<Links>
</Links>
</Error>